Wir benötigen Ihre Einwilligung zur Verwendung der einzelnen Daten, damit Sie unter anderem Informationen zu Ihren Interessen einsehen können. Klicken Sie auf "OK", um Ihre Zustimmung zu erteilen.
Standard Test Method for Carbon (Total) in Uranium Oxide Powders and Pellets By Direct Combustion-Infrared Detection Method
Automatische name übersetzung:
Standard Test Method for Carbon (Gesamt ) in Uranoxid Pulver und Pellets durch direkte Verbrennung - Infrarot- Erkennungsmethode
NORM herausgegeben am 1.6.2004
Bezeichnung normen: ASTM C1408-98(2004)
Anmerkung: UNGÜLTIG
Ausgabedatum normen: 1.6.2004
SKU: NS-11115
Zahl der Seiten: 3
Gewicht ca.: 9 g (0.02 Pfund)
Land: Amerikanische technische Norm
Kategorie: Technische Normen ASTM
Keywords:
carbon content, gadolinium oxide, gadolinium oxide-uranium oxide, impurity content, uranium oxide, ICS Number Code 27.120.30 (Fissile materials and nuclear fuel technology)
| Significance and Use | ||||||||||
|
Uranium dioxide is used as a nuclear-reactor fuel. Gadolinium oxide is used as an additive to uranium dioxide. In order to be suitable for this purpose, these materials must meet certain criteria for impurity content. This test method is designed to determine whether the carbon content meets Specifications C 753, C 776, C 888, and C 922. |
||||||||||
| 1. Scope | ||||||||||
|
1.1 This test method covers the determination of carbon in nuclear-grace uranium oxide powders and pellets to determine compliance with specifications. 1.2 Gadolinium oxide (Gd2O3) and gadolinium oxide-uranium oxide powders and pellets may also be analyzed using this test method. 1.3 This test method covers the determination of 5 to 500 µg of residual carbon. 1.4 This test method describes an induction furnace carrier gas combustion system equipped with an infrared detector. It may also be applied to a similar instrument equipped with a thermal conductivity detector. 1.5 The preferred system of units is micrograms carbon per gram of uranium (µg/g U). 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||||||||
| 2. Referenced Documents | ||||||||||
|
Wollen Sie sich sicher sein, dass Sie nur die gültigen technischen Normen verwenden?
Wir bieten Ihnen eine Lösung, die Ihnen eine Monatsübersicht über die Aktualität der von Ihnen angewandten Normen sicher stellt.
Brauchen Sie mehr Informationen? Sehen Sie sich diese Seite an.
Letzte Aktualisierung: 2026-05-23 (Zahl der Positionen: 2 279 848)
© Copyright 2026 NORMSERVIS s.r.o.